CryoArc Tera |
CryoArc Tera for Cassettes |
Dimensions (H X W X D) |
3,175 x 1,524 x 2,007mm
125.00’’ x 60.00” x 79.00’’ |
3,179 x 1,524 x 2,007mm
125.1’’ x 60.00” x 79.00’’ |
Storage capacity |
630 x 2’’ cryoboxes or equivalent |
990 x 250ml cryo cassettes |
Weight (empty) |
1,149kg
2,552lbs |
1,054kg
2,231lbs |
Weight (full) |
2,061kg
4,538lbs |
1,563kg
3,441lbs |
Maximum floor load |
665kg/m2
136lbs/ft2 |
512kg/m2
105lbs/ft2 |
Maximum point load (4-points) |
403kg
887lbs |
403kg
887lbs |
Voltage |
220V, single phase
(110V single phase) |
220V, single phase
(110V single phase) |
Amps |
2A
(1A) |
2A
(1A) |
Power output (idle) |
100 Watts |
100 Watts |
Power output (MAX) |
220 Watts |
220 Watts |
Plug type |
CEE 7/4
(NEMA 5-15) |
CEE 7/4
(NEMA 5-15) |
LN2 |
(IEC TYPE F)
(IEC TYPE B) |
(IEC TYPE F)
(IEC TYPE B) |
Connection type / size |
1/2” NPT female fitting to existing vacuum jacket line |
1/2” NPT female fitting to existing vacuum jacket line |
Supply pressure (open valve) |
1.52-2.41bar
(22-35 psig) |
1.52-2.41bar
(22-35 psig) |
Capacity (vapor phase) |
295L |
295L |
Static evaporation rate (nominal) |
13.6L/day (11.00kg/day) |
13.6L/day (11.00kg/day) |
Static holding time vapor |
21 days |
21 days |
Operating temperature |
-190°C or lower
-310°F or lower |
-190°C or lower
-310°F or lower |
Operating environment temperature |
18°C-27°C
65°F-80°F |
18°C-27°C
65°F-80°F |
Operating environment humidity |
Max 50% rH (non-condensing) |
Max 50% rH (non-condensing) |
Warranty |
1-Year; includes onsite service, travel, and spare parts. Preventative maintenance and wear parts not included. |
1-Year; includes onsite service, travel, and spare parts. Preventative maintenance and wear parts not included. |
Other |
Installation and operator training included |
Installation and operator training included |
Regulatory compliance |
China RoHS, Conflict Materials, EU REACH, EU RoHS, CE Mark Directives
[Machinery Directive, EMC Directive, and RoHS Directive] |
China RoHS, Conflict Materials, EU REACH, EU RoHS, CE Mark Directives
[Machinery Directive, EMC Directive, and RoHS Directive] |